Validity and intra-examiner reliability of the Hawk goniometer versus the universal goniometer for the measurement of range of motion of the glenohumeral joint / Sagrario Pérez-de la Cruz, Óscar Arellano de León, Néstor Pérez Mallada, Antonio Vargas Rodríguez
Material type: Continuing resourceISSN: 1350-4533Subject(s): Hawk goniometer | Universal goniometer | Range of motion | Assessment | Shoulder In: Medical Engineering and Physics -- 2021, v. 89, p. 7-11Summary: Aim: The aim of this study was to establish the validity of an external measurement system (the Hawk goniometer). This is a digital device which measures joint ranges compared to the universal goniometer for the measurement of shoulder range of motion in healthy adults. Material and methods: a correlational study with 157 healthy volunteers completed six shoulder movements (forward flexion, abduction, extension, adduction and internal and external rotation) with each shoulder. The degree of agreement between each goniometer and Hawk measurement was assessed using Intra-class Correlation Coefficients (ICC) and Bland-Altman 95% limits of agreement (LOA). Results: the tests showed a very strong relationship between the readings of both devices (CI between 0.81 and 0.99) and there were no significant differences between the mean readings of both devices. An intraclass correlation coefficient (ICC) of above 0.9 was obtained, indicating a high intra-evaluator reliability of the Hawk goniometer in repeated measurements of shoulder range of motion. Conclusions: the Hawk goniometer is a valid and reliable element for the objective measurement of the range of motion at the shoulder joint.Item type | Current library | Collection | Call number | Status | Date due | Barcode |
---|---|---|---|---|---|---|
Artículo de revista | Revistas y artículos | Non-fiction | PP (Browse shelf(Opens below)) | Available | 5706 |
PDF en biblioteca
Bibliografía: p. 10-11
Aim: The aim of this study was to establish the validity of an external measurement system (the Hawk goniometer). This is a digital device which measures joint ranges compared to the universal goniometer for the measurement of shoulder range of motion in healthy adults.
Material and methods: a correlational study with 157 healthy volunteers completed six shoulder movements (forward flexion, abduction, extension, adduction and internal and external rotation) with each shoulder. The degree of agreement between each goniometer and Hawk measurement was assessed using Intra-class Correlation Coefficients (ICC) and Bland-Altman 95% limits of agreement (LOA).
Results: the tests showed a very strong relationship between the readings of both devices (CI between 0.81 and 0.99) and there were no significant differences between the mean readings of both devices.
An intraclass correlation coefficient (ICC) of above 0.9 was obtained, indicating a high intra-evaluator reliability of the Hawk goniometer in repeated measurements of shoulder range of motion.
Conclusions: the Hawk goniometer is a valid and reliable element for the objective measurement of the range of motion at the shoulder joint.
There are no comments on this title.